Controller

Flowchart

Technical Parameters
| Name | Basic | Minimal-Pyrogen | Low Organic | Comprehensive |
| Model | Smart-RS24 | Smart-RS24UF | Smart-RS24UV | Smart-RS24UVF |
| Pure water output* | 2-stage RO water 24 L/h | |||
| Ultrapure water output | 1.2–2.0 L/min | |||
| Ultrapure (UP) water specifications | ||||
| Resistivity (25 ℃) | 18.2 MΩ.cm | |||
| Heavy metal ions | < 0.01 ppb | |||
| Total organic carbon (TOC)* | N/A | Effectively reduces TOC | ||
| Bacteria | < 0.01 cfu/mL | |||
| Particles (> 0.2 μm) | < 1/mL | |||
|
Pyrogen (endotoxin) |
N/A | < 0.001 Eu/mL | N/A | < 0.001 Eu/mL |
| Ribonuclease (RNases) | N/A | < 1 pg/mL | N/A | < 1 pg/mL |
Deoxyribonuclease (DNases) | N/A | < 5 pg/mL | N/A | < 5 pg/mL |
| RO water specifications | ||||
| 1-stage RO water | Ion rejection rate 98%–99% (new membrane), organic rejection rate > 99% (MW > 200 Daltons), particle/bacteria rejection rate > 99% | |||
| 2-stage RO water | Ion rejection rate > 99% (new membrane), conductivity ≤ 5 µS/cm, organic/particle/bacteria rejection rate > 99%; conforms to Chinese standard for Type III water quality (RRO/RS only) | |||
| Raw water requirements | Municipal drinking (tap) water; temperature: 5–35 °C; pressure: 1.0–4.0 bar | |||
| Dimensions/Weight | 410×220×420 mm (L×W×H)/approx. 18 kg | |||
| Electrical requirements | 100–240 V, 50/60 Hz | |||
| Power | 72 W | |||
| System configuration | Main unit(with 1 set of purification cartridges) + standard external 15L pressure tank + TDS/conductivity meter + accessory kit | |||
* The value is affected by water temperature, feed water quality, and total contaminant load.