实验室纯水系统

Controller

Flowchart


Technical Parameters

NameBasicMinimal-PyrogenLow OrganicComprehensive
ModelBenchmark-DBenchmark-DUFBenchmark-DUV

Benchmark-DUVF

Pure water output*Depends on feed water pressure and flow rate
Max. flow rate1.2–2.0 L/min
Ultrapure (UP) water specifications 
 Resistivity (25 ℃)10–15 MΩ.cm
 Heavy metal ions< 0.01 ppb
 Total organic carbon (TOC)*< 10 ppb< 3 ppb
 Bacteria< 0.01 cfu/mL
 Pyrogen (endotoxin)N/A< 0.001 Eu/mLN/A< 0.001 Eu/mL
 Particles (> 0.2 μm)< 1/mL
 Ribonuclease (RNases)N/A< 1 pg/mLN/A< 1 pg/mL

 Deoxyribonuclease (DNases)

N/A< 5 pg/mLN/A< 5 pg/mL
EDI water specifications 
 Resistivity (25 ℃)> 10 MΩ.cm
 Silicon rejection rate> 99.9%
 
Raw water requirementsRO water, distilled water, and pure water of higher grades; temperature: 5–35℃; pressure: 0–1 bar
Dimensions/Weight385×500×650 mm (L×W×H)/approx. 25 kg
Electrical requirements100–240 V, 50/60 Hz
Power120 W
System configurationMain unit (with 1 set of purification cartridges) + built-in 12L clean water tank (with level indicator) + accessory kit

* The value is affected by water temperature, feed water quality, and total contaminant load.